Specht vacuum regulator - rail-mounted units
COMPRESSED-AIR OPERATED
Medical vacuum regulator ejector is based on a compressed air ejector for regulation of suction;
with vacuum meter, fine control valve and quick action stop valve;
Rail unit incl. rail clamp for attaching to a rail system for supporting medical equipment in line with DINEN 12218 (25x10mm or 30x10mm);
Air inlet integrated into the rail clamp (NIST). With the fine control valve it is possible to regulate the vacuum infinitely.
All Specht regulators meet DIN EN ISO 10079-3 specifications.
- TECHNICAL DATA
Design: Vacuum generation using the ejector principle. Brass, chrome-plated. Medium: Medical compressed air (AIR) Inlet: NIST – gas-specific in accordance with DIN EN 18082 Primary pressure: 450 kPa ± 100 kPa (in accordance with DIN EN ISO 7396-1) Suction inlet: 9/16“-18 UNF with vacuum nozzle Dimensions (WxHxD)/ weight: 110 x 160 x 115 mm / 1070 g (Specht -90)
110 x 160 x 115 mm / 1070 g (Specht -60)
120 x 160 x 115 mm / 1170 g (Specht -16)
120 x 160 x 115 mm / 1170 g (Specht -10) - PERFORMANCE PARAMETERS
Suction adjustable between: Suction flowrate Compressed air consumption Specht -90 0 and -90 kPa > 20 l/min at -90 kPa max. 30 l/min at -90 kPa Specht -60 0 and -60 kPa > 25l/min at -60 kPa max. 30 l/min at -60 kPa Specht -16 0 and -16 kPa > 18 l/min at -16 kPa max. 20 l/min at -16 kPa Specht -10 0 and -10 kPa > 18 l/min at -10 kPa max. 15 l/min at -10 kPa - Order numbers
SPECHT VACUUM REGULATOR, AIR, RAIL-MOUNTED UNIT, NIST, WITHOUT CONNECTING HOSE*
Specht -90 vacuum regulator, AIR, 0 to -90 kPa, rail-mounted unit, NIST 902.625 Specht -60 vacuum regulator, AIR, to -60 kPa, rail-mounted unit, NIST 901.592 Specht -16 vacuum regulator, AIR, 0 to -16 kPa,rail-mounted unit, NIST 902.626 Specht -10 vacuum regulator, AIR, 0 to -10 kPa, rail-mounted unit, NIST 902.672
*For connecting hose, see product group accessories