Hawk pressure regulator unit
WITH BOTTLE CONNECTOR
Pressure regulator units for connection to an oxygen cylinder, with volume
manometer and safety valve.
Includes a tube flowmeter for dosing the flow and an ejector with a vacuum
gauge and fine-dosing valve.
The outlet has a 9/16“ thread for direct connection
The following accessories can be attached:
• Drainage system (recommended)
• Nebuliser or humidifier (disposable or reusable)
- Technical Data
Pressure regulator unit
Design: Spring-loaded membrane pressure regulator
in accordance with DIN EN ISO 10524-1
Brass, chrome-platedMedium: O2 - oxygen Inlet Gas-specific hand connector in accordance with DIN 477-1 Primary pressure: max. 20.000 kPa (P1) Outlet pressure: 450 kPa ± 50 kPa (P2) Manometer: 0-31,500 kPa display Flowrate: max. 80 l/min Flowmeter
Design: Röhren-Flowmeter gemäß DIN EN ISO 15002, druckkompensiert, inkl. Feinregulierventil Material: Brass, chrome-plated Outlet: 9/16" - 18 UNF thread Flowrate: 0 – 15 l/min Accuracy: ± 10 % full scale Ejector for vacuum generation
Design: Vacuum generation using the ejector principle,
in accordance with DIN EN ISO 10079-3Material: Brass, chrome-plated Outlet: 9/16“-18 UNF with vacuum nozzle 6mm Dimensions/weight
(BxHxT):160 x 190 x 180 mm / ca. 1600 g Performance parameters: Adjustable between 0 and -90 kPa Consumption: max. 30 l/min at -90 kPa Suction flowrate: min. 20 l/min at -90 kPa (± 10%) - Order numbers
Hawk O2: Suction + tube flowmeter without humidifier,
VAC: 0 to -90 kPa, flowrate: 0-15 l/min910.003